• Six contacts for Hall effect, four contacts for van der Pauw measurement
• Sample size: max 10mm x 10mm
• up to 12 contacts to sample
• 29mm OD sample insert (other sizes are optional)
• 25 spare sample holders
• Measurement electronics and control software
• Temperature range: 1-300K
• Rotation: 00-3600 with 0.016 resolution
• Sample size: max 5mm x 10mm
• up to 12 contacts to sample
• 29mm OD sample insert (other sizes are optional)
• 25 spare sample holders
• Measurement electronics and control software
• Temperature range: 1—300K
• KF50 or KF40 flange option
• Source current programming range : 50pA to IA •
• Voltage measurement range: 100mV (100nV resolution) to 100V (IOpV resolution)
• Temperature range: 1-300K
AC Resistance / Hall Effect Measurements
• Source current programming range: 2nA to 100mA *
• Frequency range: IHz to 1kHz
• Real and imaginary parts of impedance Z(T)
• Voltage noise (with ow noise voltage preamplifier): lnVNHz noise floor at 1kHz at 60dB gain
• Sample Size: Max 5 mm diameter
• AC Frequency Range: IOHz to 10kHz
• AC Field Amplitude Range: 2mOe to 150e
• Sensitivity: 2 x 10-7 emu (2 x 10 IOAm2at 10kHz)
• Real and imaginary parts of susceptibility (T)
• Includes calibrated sensor at sample stage/pick-up coils
• Measurement electronics and control software
• Temperature Range: 1-300K
• Low noise voltage preamplifier for AC Resistance
. 1 5 nV/QHz voltage noise floor at 1 kHz
Voltage input range: ± 5V at xl gain
Common mode rejection: 120dB at 1 kHz
• Calibrated Thermometer and Heater at the Sample Stage
• 400K High Temperature Option for measurement probes